GMOC

The GMOC (Gersh Monthly Output Check) Phantom is designed to streamline the entire Monthly Output Check Process. Ask about pairing with our iQA Auto Analysis Software!
GMOC is hosted in the electron cage for fast and simple routines

Key Features

  • The Lightweight GMOC phantom fits securely in LINAC electron collimator to ensure accurate repeatability
  • Therapist-Friendly Setup - No phantom alignment or couch shifts required
  • Dose Calibration / Output Constancy
  • Beam Energy Constancy
  • Rotational Dose Output Constancy
  • kV / MV Image System Alignment
  • Isocentric Room Laser Alignment
  • OBI Software Image Distance Measuring Accuracy
  • Optical Distance Indicator Accuracy
  • Optical Reticle Alignment
  • Light Field-Size Accuracy

Full Description

The IMT GMOC phantom, developed with our clinical partners, was designed with simplicity and ergonomics in mind while providing a broad range of Monthly Output testing functions. This unique phantom is lightweight, reproducible and efficient.

The GMOC fulfills a multitude of tests performed during a clinic’s typical Monthly Output Checks providing complete verification of LINAC setup, alignment, constancy, and accuracy.

The GMOC utilizes an ion chamber located at the central axis of the phantom (and thus the central axis of the LINAC), and is configured to your clinic’s specificaitons. In order to ensure proper and repeatable alignment to the LINAC, an aluminum mounting plate is configured to your LINAC specification.